FDA Adverse Event Death Summary report: N

RESPIRATORY HUMIDIFIER

MDR report key: 1002627 · Received February 27, 2008

Report

Report Number
9611451-2008-00069
Event Type
Death
Date Received
February 27, 2008
Date of Event
January 27, 2008
Report Date
January 28, 2008
Manufacturer
FISHER & PAYKEL HEALTHCARE, LTD.
Product Code
BTT
PMA / PMN Number
K983112
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE HAS BEEN QUARANTINED BY THE HOSP. RESULTS: RESPIRONICS ESPRIT VENTILATOR. CONCLUSIONS: THE HOSP CONTINUED TO USE THE RT110 ADULT BREATHING CIRCUIT WITH THE RESPIRONICS ESPRIT VENTILATOR CONTRARY TO FISHER & PAYKEL HEALTHCARE RECOMMENDATIONS. WE ARE CONTINUING TO INVESTIGATE THIS REPORTED EVENT AND SEEK ADD'L INFO FROM THE REPORTING HOSP.

Description of Event or Problem · 1

A HOSP REPORTED TO OUR DIST THAT "ONE PT CODED AND EVENTUALLY EXPIRED" WHILST ON A RESPIRONICS ESPRIT VENTILATOR, FISHER & PAYKEL HEALTHCARE MR850 RESPIRATORY HUMIDIFIER AND THE RT110 ADULT BREATHING CIRCUIT. F/U INFO FROM THE HOSP INCLUDED: THE HOSP BEGAN USING DUAL HEATED BREATHING CIRCUITS "IN 2007, AND THEY HAD BEEN EXPERIENCING CONDENSATE ISSUES IN THEIR VENTILATOR SYSTEMS. IN 2008, A MALE PT "CODED" AND THE HOSP ATTEMPTED TO RESUSCITATE THE PT FOR APPROX 30 MINS "WITHOUT SUCCESS". THE "VENTILATOR DID ALARM" AND THE "NURSE THAT DISCOVERED IT SAW 'OCCLUSION' IN THE WINDOW" OF THE VENTILATOR. THE HOSP IS CURRENTLY IN POSSESSION OF THE VENTILATOR AND HUMIDIFIER. THE PRIOR STATUS OF THE PT IS UNK. IT SHOULD BE NOTED THAT THE HOSP CONTINUED TO USE THE RT110 ADULT BREATHING CIRCUIT WITH THE RESPIRONICS ESPRIT VENTILATOR CONTRARY TO FISHER & PAYKEL HEALTHCARE RECOMMENDATIONS. THE VENTILATOR MFR HAS BEEN NOTIFIED OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 RESPIRATORY HUMIDIFIER BTT FISHER & PAYKEL HEALTHCARE, LTD. MR850

Patients

Seq Age Sex Outcome Treatment
1 YR Death RESPIRONICS ESPRIT VENTILATOR| BREATHING CIRCUIT| FISHER & PAYKEL HEALTHCARE RT110 ADULT