FDA Adverse Event Malfunction Summary report: N

SYRINGE 0.3ML 29GA 1/2IN 7BAG 420CAS JP

MDR report key: 10022604 · Received May 4, 2020

Report

Report Number
1920898-2020-00470
Event Type
Malfunction
Date Received
May 4, 2020
Date of Event
April 14, 2020
Report Date
May 6, 2020
Manufacturer
BD MEDICAL - DIABETES CARE
Product Code
FMF
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION: 2 SAMPLES WERE RETURNED AND PICTURES WERE FORWARDED TO THE MANUFACTURING FACILITY OF 3/10CC SYRINGES. CUSTOMER STATES THAT THE HUB WAS DETACHED WITH THE SHIELD. THE ATTACHED PHOTOS WERE EXAMINED AND EXHIBITED THE HUB-NEEDLE/SHIELD ASSEMBLY SEPARATED FROM THE BARREL. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR BATCH# 9231316. ALL INSPECTIONS AND CHALLENGES WERE PERFORMED PER THE APPLICABLE OPERATIONS QC SPECIFICATIONS. THERE WAS ONE (1) NOTIFICATION [200842714] NOTED THAT DID NOT PERTAIN TO THE COMPLAINT. PROCESS SUMMARY: AUTOMATIC SYRINGE ASSEMBLY MACHINE, WHICH FEEDS 3/10CC, SYRINGE COMPONENTS (BARREL, STOPPER, PLUNGER, NEEDLE ASSEMBLY & CAP) AND ASSEMBLES THESE COMPONENTS. THIS MACHINE CONSISTS OF A BARREL CLEANING DIAL, LUBRICATION DIAL, PLUNGER/STOPPER ASSEMBLY DIAL, SYRINGE ASSEMBLY DIAL, AND VARIOUS INSPECTIONS AND TRANSFER DIALS. DHR, L2L DISPATCHES, AND LOGBOOK ENTRIES WERE LOOKED AT, NOTHING PERTAINING TO THIS DEFECT WAS FOUND. ROOT CAUSE FOR THIS DEFECT CANNOT BE DETERMINED. CAPA#1122103 WAS INITIATED. H3 OTHER TEXT : SEE H.10.

Description of Event or Problem · 0

IT WAS REPORTED THAT SYRINGE 0.3ML 29GA 1/2IN 7BAG 420CAS JP HUB DETACHED BEFORE USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: A HUB WAS DETACHED WITH THE SHIELD.

Additional Manufacturer Narrative · 1

A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT SYRINGE 0.3ML 29GA 1/2IN 7BAG 420CAS JP HUB DETACHED BEFORE USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: A HUB WAS DETACHED WITH THE SHIELD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
483556 SYRINGE 0.3ML 29GA 1/2IN 7BAG 420CAS JP SYRINGE FMF BD MEDICAL - DIABETES CARE 9231316

Patients

Seq Age Sex Outcome Treatment
1 Other