FDA Adverse Event Malfunction Summary report: N

PANALOK LOOP RC ANCHOR

MDR report key: 1001864 · Received February 25, 2008

Report

Report Number
1221934-2008-00094
Event Type
Malfunction
Date Received
February 25, 2008
Date of Event
February 15, 2008
Report Date
February 15, 2008
Manufacturer
DEPUY MITEK
Product Code
MBI
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

NOTHING IS BEING RETURNED FOR A FAILURE ANALYSIS, HOWEVER, THIS TYPE OF FAILURE HAS HISTORICALLY BEEN ATTRIBUTED TO USER TECHNIQUE. FROM AN ENGINEERING PERSPECTIVE, DAMAGE TO THE INSERTER, "TIP BREAKAGE", MAY HAVE BEEN CAUSED BY OFF-ANGLE OR OFF AXIS INSERTION INTO THE BONE HOLE. OFF ANGLE/AXIS INSERTION IS THE MOST LIKELY CAUSE FOR THE INSERTER DISTAL TIP FAILURE, USUALLY THE RESULT OF BENDING AND/OR TWISTING THE ANCHOR DURING DEPLOYMENT DUE TO ANCHOR/BONE HOLE MISALIGNMENT. THE IFU STATES NOT TO TWIST AND/OR BEND THE INSERTER UPON INSERTION AS THE ANCHOR, SUTURE AND/OR INSERTER TIP MAY BE DAMAGED. NO FURTHER ACTION IS WARRANTED, HOWEVER, MITEK WILL CONTINUE TO TRACK ANY RELATED COMPLAINTS WITHIN THIS DEVICE FAMILY AS A MEANS OF MONITORING THE EXTENT WITH WHICH THIS COMPLAINT IS OBSERVED IN THE FIELD.

Description of Event or Problem · 1

OUR AFFILIATE REPORTS THAT DURING AN ARTHROSCOPIC SHOULDER REPAIR, A PORTION OF THE DISTAL TIP OF A PANALOK INSERTER BROKE OFF INTO ITS RESPECTIVE ANCHOR AND REMAINS WITHIN THE ANCHOR WHICH REMAINS SEATED IN THE BONE HOLE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT FURTHER INCIDENT OR HARM TO THE PT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PANALOK LOOP RC ANCHOR SOFT TISSUE FIXATION DEVICE MBI DEPUY MITEK 210702 3077426

Patients

Seq Age Sex Outcome Treatment
1 UNK