PANALOK LOOP RC ANCHOR
Report
- Report Number
- 1221934-2008-00094
- Event Type
- Malfunction
- Date Received
- February 25, 2008
- Date of Event
- February 15, 2008
- Report Date
- February 15, 2008
- Manufacturer
- DEPUY MITEK
- Product Code
- MBI
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
NOTHING IS BEING RETURNED FOR A FAILURE ANALYSIS, HOWEVER, THIS TYPE OF FAILURE HAS HISTORICALLY BEEN ATTRIBUTED TO USER TECHNIQUE. FROM AN ENGINEERING PERSPECTIVE, DAMAGE TO THE INSERTER, "TIP BREAKAGE", MAY HAVE BEEN CAUSED BY OFF-ANGLE OR OFF AXIS INSERTION INTO THE BONE HOLE. OFF ANGLE/AXIS INSERTION IS THE MOST LIKELY CAUSE FOR THE INSERTER DISTAL TIP FAILURE, USUALLY THE RESULT OF BENDING AND/OR TWISTING THE ANCHOR DURING DEPLOYMENT DUE TO ANCHOR/BONE HOLE MISALIGNMENT. THE IFU STATES NOT TO TWIST AND/OR BEND THE INSERTER UPON INSERTION AS THE ANCHOR, SUTURE AND/OR INSERTER TIP MAY BE DAMAGED. NO FURTHER ACTION IS WARRANTED, HOWEVER, MITEK WILL CONTINUE TO TRACK ANY RELATED COMPLAINTS WITHIN THIS DEVICE FAMILY AS A MEANS OF MONITORING THE EXTENT WITH WHICH THIS COMPLAINT IS OBSERVED IN THE FIELD.
OUR AFFILIATE REPORTS THAT DURING AN ARTHROSCOPIC SHOULDER REPAIR, A PORTION OF THE DISTAL TIP OF A PANALOK INSERTER BROKE OFF INTO ITS RESPECTIVE ANCHOR AND REMAINS WITHIN THE ANCHOR WHICH REMAINS SEATED IN THE BONE HOLE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT FURTHER INCIDENT OR HARM TO THE PT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PANALOK LOOP RC ANCHOR | SOFT TISSUE FIXATION DEVICE | MBI | DEPUY MITEK | 210702 | 3077426 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |