FDA Adverse Event
Malfunction
Summary report: N
MAVERICK2 MONORAIL PTCA CATHETER
MDR report key: 1001745
·
Received February 19, 2008
Report
- Report Number
- 2134265-2008-00461
- Event Type
- Malfunction
- Date Received
- February 19, 2008
- Date of Event
- January 29, 2008
- Report Date
- January 30, 2008
- Manufacturer
- BOSTON SCIENTIFIC
- Product Code
- LOX
- PMA / PMN Number
- P860019/S179
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVAL: THE COMPLAINANT INDICATED THAT THE DEVICE WILL NOT BE RETURNED FOR EVAL; THEREFORE, A FAILURE ANALYSIS OF THE COMPLAINT DEVICE COULD NOT BE COMPLETED. A REVIEW OF THE BATCH HISTORY, HISTORICAL TRENDING, AND SIMILAR COMPLAINT TRENDING REVIEW FOR THE PRODUCT FAMILY WILL CONDUCTED. IF THERE IS ANY FURTHER RELEVANT INFO FROM THAT REVIEW, A SUPPLEMENTAL MEDWATCH WILL BE FILED.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING A PERCUTANEOUS CORONARY INTERVENTION (PCI) PROCEDURE, A BALLOON RUPTURE OCCURRED. THE PHYSICIAN ADVANCED THE 1.5X15MM MAVERICK2 BALLOON TO THE LESION AND INFLATED IT TO 10 ATMS AND IT RUPTURED. THERE WERE NO PT COMPLICATIONS. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH A DIFFERENT DEVICE. PT STATUS IS REPORTED IS "GOOD".
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MAVERICK2 MONORAIL PTCA CATHETER | LOX CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS | LOX | BOSTON SCIENTIFIC | MAVERICK 15 / 1.5 | 9004304 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |