TAXUS EXPRESS2 PACLITAXEL-ELUTING CORONARY STENT SYSTEM
Report
- Report Number
- 2134265-2008-00473
- Event Type
- Malfunction
- Date Received
- February 19, 2008
- Date of Event
- January 28, 2008
- Report Date
- January 28, 2008
- Manufacturer
- BOSTON SCIENTIFIC
- Product Code
- NIQ
- PMA / PMN Number
- P030025
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- PHYSICIAN
Narratives
THE COMPLAINANT INDICATED THAT THE DEVICE WILL NOT BE RETURNED FOR EVALUATION; THEREFORE, A FAILURE ANALYSIS OF THE COMPLAINT DEVICE COULD NOT BE COMPLETED. A REVIEW OF THE BATCH HISTORY, HISTORICAL TRENDING, AND SIMILAR COMPLAINT TRENDING REVIEW FOR THE PRODUCT FAMILY WILL BE CONDUCTED. IF THERE IS ANY FURTHER RELEVANT INFORMATION FROM THAT REVIEW, A SUPPLEMENTAL MEDWATCH WILL BE FILED.
IT WAS REPORTED THAT DURING A CORONARY ARTERY DRUG ELUTING STENTING TREATMENT PROCEDURE, STENT DAMAGE OCCURRED. THE 99% STENOSED TARGET LESION WAS LOCATED IN THE SEVERELY TORTUOUS AND CALCIFIED MID LEFT ANTERIOR DESCENDING (LAD) ARTERY. WHEN THE PHYSICIAN ATTEMPTED TO ADVANCE THE 2.50X12MM TAXUS EXPRESS2 DRUG ELUTING STENT (DES), THERE WAS DIFFICULTY ADVANCING TO THE LESION. WHEN THE STENT DELIVERY SYSTEM (SDS) WAS WITHDRAWN FROM INSIDE THE PATIENT, THE PHYSICIAN NOTICED THAT THE PROXIMAL EDGE OF THE STENT WAS LIFTED UP. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE WITH NO PATIENT COMPLICATIONS REPORTED. THE PATIENT'S CURRENT CONDITION IS LISTED AS "GOOD".
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TAXUS EXPRESS2 PACLITAXEL-ELUTING CORONARY STENT SYSTEM | NIQ STENT, CORONARY, DRUG-ELUTING | NIQ | BOSTON SCIENTIFIC | 2.50X12MM | 9550168 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |