STRATA II VALVE, REGULAR, INTEGRAL SNAP
MDR report key: 1001647
·
Received February 18, 2008
Report
- Report Number
- 2021898-2008-00044
- Event Type
- Malfunction
- Date Received
- February 18, 2008
- Date of Event
- January 18, 2008
- Report Date
- January 25, 2008
- Manufacturer
- MEDTRONIC NEUROSURGERY
- Product Code
- JXG
- PMA / PMN Number
- K042465
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE RETURNED VALVE WAS NOT PATENT DUE TO A LARGE JAGGED TEAR IN THE TOP OF THE RESERVOIR. THIS DAMAGE PRECLUDED ALL FURTHER TESTING. A REVIEW OF THE MFG RECORDS SHOWED NO ANOMALIES. ALL OF OUR PRODUCTS ARE 100% TESTED AT THE TIME OF MFR. THE INSTRUCTIONS FOR USE THAT ACCOMPANIES THE VALVES CAUTION THAT CARE SHOULD BE TAKEN IN HANDLING THE VALVES AS SILICONE HAS A LOW CUT AND TEAR RESISTANCE.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE VALVE HAD A HOLE IN IT.