FDA Adverse Event Malfunction Summary report: N

TUBE RNA PLH 16X100 2.5 PLBLCE CLR

MDR report key: 10015736 · Received April 30, 2020

Report

Report Number
9617032-2020-00382
Event Type
Malfunction
Date Received
April 30, 2020
Date of Event
April 13, 2020
Report Date
June 11, 2020
Manufacturer
BECTON, DICKINSON AND COMPANY (BD)
Product Code
JKA
UDI-DI
54053727621653
PMA / PMN Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION SUMMARY BD HAD NOT RECEIVED SAMPLES OR PHOTOS FROM THE CUSTOMER FACILITY FOR EVALUATION. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR THE INCIDENT LOT AND, BASED ON THIS REVIEW, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET; THERE WERE NO RELATED QUALITY NON-CONFORMANCES DURING MANUFACTURING OF THE PRODUCT.

Description of Event or Problem · 0

IT WAS REPORTED BEFORE USE THE TUBE RNA PLH 16X100 2.5 PLBLCE CLR TUBES (WITHOUT SAMPLE) STORED AT THE WRONG TEMPERATURE AND CRACKED TUBE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "LAB RECEIVED 78 FROZEN PAXGENE BLOOD RNA TUBES FROM ITALY AND 35 OF THEM WERE CRACKED ON THE BOTTOM. THE DAMAGE TUBES WERE NOTICED WHEN THE SAMPLES WERE PULLED FROM THE -80OC FREEZER.

Additional Manufacturer Narrative · 1

MEDICAL DEVICE EXPIRATION DATE: UNKNOWN. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. DEVICE MANUFACTURE DATE: UNKNOWN.

Description of Event or Problem · 1

IT WAS REPORTED BEFORE USE THE TUBE RNA PLH 16X100 2.5 PLBLCE CLR TUBES (WITHOUT SAMPLE) STORED AT THE WRONG TEMPERATURE AND CRACKED TUBE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "LAB RECEIVED 78 FROZEN PAXGENE BLOOD RNA TUBES FROM (B)(4) AND 35 OF THEM WERE CRACKED ON THE BOTTOM. THE DAMAGE TUBES WERE NOTICED WHEN THE SAMPLES WERE PULLED FROM THE -80OC FREEZER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
476343 TUBE RNA PLH 16X100 2.5 PLBLCE CLR BLOOD SPECIMEN COLLECTION DEVICE JKA BECTON, DICKINSON AND COMPANY (BD) 762165 UNKNOWN 54053727621653

Patients

Seq Age Sex Outcome Treatment
1 Other