FDA Adverse Event Malfunction Summary report: N

TRIGGER HANDLE F/CABLE CUTTER

MDR report key: 10015180 · Received April 30, 2020

Report

Report Number
2939274-2020-02152
Event Type
Malfunction
Date Received
April 30, 2020
Report Date
April 15, 2020
Manufacturer
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
Product Code
HTZ
UDI-DI
10886982076700
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. ADDITIONAL NARRATIVE: INVESTIGATION SUMMARY: INVESTIGATION FLOW: DAMAGE. VISUAL INSPECTION: THE TRIGGER HANDLE F/CABLE CUTTER (P/N: 03.221.007, LOT #: T162927) WAS RETURNED AND RECEIVED AT US CQ. UPON VISUAL INSPECTION, THE DEVICE WAS MISSING THE M4 SHOULDER NUT AND BOLT. NO OTHER ISSUES WERE IDENTIFIED WITH THE RETURNED COMPONENTS OF THE DEVICE. DEVICE FAILURE/DEFECT IDENTIFIED? YES. DIMENSIONAL INSPECTION: THERE WAS CONCLUSIVE EVIDENCE THAT THE RETURNED DEVICE WAS MISSING COMPONENTS, SO THE DIMENSIONAL INSPECTION WAS NOT PERFORMED. DOCUMENT/SPECIFICATION REVIEW: BASED ON THE DATE OF MANUFACTURE, THE CURRENT AND MANUFACTURED REVISION OF DRAWINGS WERE REVIEWED . COMPLAINT CONFIRMED? YES, THE DEVICE RECEIVED WAS MISSING COMPONENTS. HENCE CONFIRMING THE ALLEGATION. INVESTIGATION CONCLUSION : THE COMPLAINT CONDITION IS CONFIRMED FOR THE TRIGGER HANDLE F/CABLE CUTTER (P/N: 03.221.007, LOT #: T162927). THERE IS NO INDICATION THAT A DESIGN OR MANUFACTURING ISSUE HAS CAUSED THE COMPLAINT CONDITION AND HENCE THE ROOT CAUSE CANNOT BE DETERMINED. THE POTENTIAL CAUSE COULD BE DUE TO UNINTENDED FORCES APPLIED TO THE DEVICE. BASED ON THE INVESTIGATION FINDINGS, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST-MARKET SAFETY SURVEILLANCE ACTIVITIES. DEVICE HISTORY LOT. PART NUMBER: 03.221.007, LOT NUMBER: T162927, MANUFACTURING SITE: TUTTLINGEN, RELEASE TO WAREHOUSE DATE: (B)(6) 2018. A REVIEW OF THE DEVICE HISTORY RECORDS WAS PERFORMED FOR THE FINISHED DEVICE LOT NUMBER, AND NO NON-CONFORMANCES WERE IDENTIFIED. THE RAW MATERIAL CERTIFICATE WAS REVIEWED AND THE USED MATERIAL WAS ACCORDING TO THE SPECIFICATION OF THE DEVICE. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 0

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. H3, H6: THE DEVICE WAS RECEIVED, THE INVESTIGATION IS IN PROGRESS, NO CONCLUSION COULD BE DRAWN AT THE TIME OF FILING THIS REPORT. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 0

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. H11 : CORRECTED DATA H6: MANUFACTURING DATE UPDATED DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 0

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF THE INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 0

THERE WAS NO PATIENT INVOLVEMENT.

Additional Manufacturer Narrative · 1

ADDITIONAL NARRATIVE: REPORTER IS COMPANY REPRESENTATIVE. WITHOUT A LOT NUMBER, THE DEVICE HISTORY RECORDS REVIEW COULD NOT BE COMPLETED AS NO PRODUCT WAS RECEIVED. THE INVESTIGATION COULD NOT BE COMPLETED, NO PRODUCT WAS RECEIVED; NO CONCLUSION COULD BE DRAWN AT THE TIME OF FILING THIS REPORT. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

IT WAS REPORTED ON AN UNKNOWN DATE, ONE (1) TRIGGER HANDLE FOR CABLE CUTTER, ONE (1) INTER-LOCK SCREWDRIVER, ONE (1) DEPTH GAUGE, ONE (1) DRIVE ADAPTOR, THREE (3) REDUCTION FORCEPS AND ONE (1) POINTED REDUCTION CLAMP WERE HANDED BY THE FACILITY ORTHO DIRECTOR TO THE SALES CONSULTANT AS ALL THE DEVICES WERE BROKEN. IT IS UNKNOWN IF THERE WERE PATIENT AND SURGICAL INVOLVEMENT. THIS COMPLAINT INVOLVES EIGHT (8) DEVICES. THIS IS REPORT 1 OF 8 FOR (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
476900 TRIGGER HANDLE F/CABLE CUTTER INSTRUMENT,CUTTING,ORTHOPAEDIC HTZ WRIGHTS LANE SYNTHES USA PRODUCTS LLC T162927 10886982076700

Patients

Seq Age Sex Outcome Treatment
1 DEPTH GAUGE FOR 2.0MM AND 2.4MM SCREWS-70MM| DRIVE ADAPTOR WITH QC FOR 5.0MM SCHANZ SCREWS| INTER-LOCK SCREWDRIVER T25/3.5MM HEX/224MM| REDUCTION FORCEPS WITH POINTS NARROW-RATCHET 132MM| TRIGGER HANDLE F/CABLE CUTTER| UNK - FORCEPS