Report

Report Number
9611451-2008-00068
Event Type
Malfunction
Date Received
February 23, 2008
Report Date
January 24, 2008
Manufacturer
FISHER & PAYKEL HEALTHCARE, LTD.
Product Code
BTT
PMA / PMN Number
K983112
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS IS A MALFUNCTION THAT HAS BEEN REPORTED TO FISHER AND PAYKEL HEALTHCARE BY A HOSPITAL. THE PRODUCT IS NOT SOLD IN THE USA BUT IT IS SIMILAR TO A PRODUCT WHICH IS SOLD IN THE USA. ADD'L LOT NUMBER IS 071012, MFR DATE 10/12/2007. THE RESISTANCE OF THE HEATER WIRE ON BOTH THE INSPIRATORY AND EXPIRATORY LIMBS OF ALL 7 RETURNED CIRCUITS WERE MEASURED. RESULTS: THE RESULTS OF THESE TESTS SHOW THAT EITHER THE INSPIRATORY OR EXPIRATORY HEATER WIRE RESISTANCE OR SOMETIMES BOTH WERE OUTSIDE THE SPECIFICATION, ALWAYS ON THE HIGH SIDE. HIGHER RESISTANCES ARE USUALLY ASSOCIATED WITH IMPROPER PINNING OF THE HEATER WIRES. OUR RECORDS INDICATE THAT ONE OTHER COMPLAINT (TOTAL TWO COMPLAINTS) OF THIS NATURE HAS BEEN RECEIVED FOR THE GIVEN LOT NUMBER (0710102). CONCLUSIONS: THE CIRCUITS ARE OUT OF SPECIFICATION. THERE HAVE BEEN CHANGES MADE TO THE PINNING PROCESS RECENTLY THROUGH THE REPLACEMENT OF ALL THE PINNING MACHINES ON THE PRODUCTION LINE. THE EFFECTIVE DATE FOR THIS IMPROVEMENT IS FOR ALL PRODUCT MANUFACTURED AFTER 11/29/2007. ALL OF THE ABOVE PRODUCT WAS MANUFACTURED BEFORE THE DATE OF THE CORRECTIVE ACTION. OUR MONITORING AND TRENDING FOR THIS TYPE OF EVENT HAS A RATE OF OCCURRENCE WORLDWIDE FOR THE LAST YEAR OF 0.0022%.

Description of Event or Problem · 1

A HOSPITAL REPORTED TO OUR DISTRIBUTOR THAT 7 RT200 ADULT BREATHING CIRCUITS HAD A HEATER WIRE RESISTANCE VALUE THAT IS NOT NORMAL. NO PT CONSEQUENCE WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 NONE BTT FISHER & PAYKEL HEALTHCARE, LTD. RT200 071009

Patients

Seq Age Sex Outcome Treatment
1