FDA Adverse Event Malfunction Summary report: N

GE OEC 9800

MDR report key: 1001352 · Received February 22, 2008

Report

Report Number
1720753-2008-16346
Event Type
Malfunction
Date Received
February 22, 2008
Date of Event
February 14, 2008
Report Date
February 21, 2008
Manufacturer
GE OEC MEDICAL SYSTEMS INC
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE REPRESENTATIVE EVALUATED THE SYSTEM AND REPLACED THE GENERATOR INTERFACE PCB, FLUORO FUNCTIONS PCB, AND FOOT SWITCH. SYSTEM OPERATES AS INTENDED.

Description of Event or Problem · 1

CUSTOMER REPORTED THAT THE SYSTEM WOULD QUIT DISPLAYING IMAGES IF MOVED DURING FLUORO. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 GE OEC 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS INC 9800 NA

Patients

Seq Age Sex Outcome Treatment
1