FDA Adverse Event Malfunction Summary report: N

INRATIO

MDR report key: 1001331 · Received February 19, 2008

Report

Report Number
2954730-2008-00071
Event Type
Malfunction
Date Received
February 19, 2008
Date of Event
January 29, 2008
Report Date
February 14, 2008
Manufacturer
HEMOSENSE SJ
Product Code
JPA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

INRATIO PRECISION DATA PROVIDED BY END-USER LOT: 070274. FIRST TEST INR = 1.3; SECOND TEST INR = 1.8; THIRD TEST INR = 4.4; MEAN = 2.5; SD = 1.7; %CV = 66.57. THE %CV IS GREATER THAN 20%. PER INTERNAL PROCEDURE, THE PRECISION FAILS THE CRITERIA FOR PRECISION. THE TEST IS CONSIDERED PRECISE AND FURTHER TESTING IS REQUIRED AT THIS TIME. DISCREPANT RESULTS (ACCURACY) COMPARISON OF INRATIO TEST WITH LAB RESULTS PROVIDED BY END-USER AT TIME COMPLAINT WAS FILED: DATE: 2008, INRATIO: 4.0, COAGUCHECK: 4.7, MEAN: 4.4, CONFIDENCE LIMITS: 2.5 - 6.5. PER INTERNAL PROCEDURE, THE MEAN OF THE INRATIO METER AND COMPARATIVE SYS INR WERE CALCULATED. BOTH INRATIO AND LAB VALUES ARE WITHIN THE CONFIDENCE LIMITS FOR INR TESTING. THE RESULTS ARE NOT CONSIDERED DISCREPANT WITHIN THE CONTEXT OF THE DOCUMENTED VARIABILITY FOR INR TESTING. THEREFORE, FURTHER TESTING IS NOT REQUIRED AT THIS TIME.

Description of Event or Problem · 1

CALLER ALLEGES IMPRECISION WITH INRATIO. RESULTS AS FOLLOWS: FIRST TEST INR = 1.3; SECOND TEST INR = 1.8; THIRD TEST INR = 4.4. CALLER ALLEGES INACCURACY WITH INRATIO. RESULTS AS FOLLOWS: DATE: 2008, INRATIO: 4.0, COAGUCHECK: 4.7.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INRATIO PROTHROMBIN TIME TEST JPA HEMOSENSE SJ BOXED STRIPS, 12-PK MULTILANGUAGE 070274

Patients

Seq Age Sex Outcome Treatment
1