FDA Adverse Event Injury Summary report: N

MENTOR MEMORYGEL BREAST IMPLANT

MDR report key: 10012225 · Received April 29, 2020

Report

Report Number
1645337-2020-05963
Event Type
Injury
Date Received
April 29, 2020
Date of Event
March 11, 2020
Report Date
April 6, 2020
Manufacturer
MENTOR TEXAS
Product Code
FTR
UDI-DI
00081317000273
PMA / PMN Number
P030053
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

ON (B)(6) 2020, THE PRODUCT INVESTIGATION WAS COMPLETED. DEVICE INVESTIGATION SUMMARY: DURING THE VISUAL INSPECTION, THE DEVICE WAS FOUND TO BE RUPTURED. ADDITIONALLY, AN ANOMALY WAS OBSERVED IN THE EDGES OF THE RUPTURE, CONSISTENT WITH A CREASE/FOLD. THE EVALUATION DETERMINED THAT THE LOSS OF SHELL INTEGRITY IS CONSISTENT WITH A CREASE FOLD RUPTURE OF THE IMPLANT SHELL, WHICH IS A KNOWN INHERENT RISK OF GEL-FILLED MAMMARY PROSTHESIS. RUPTURE CAN OCCUR AT ANY TIME AFTER IMPLANTATION, BUT IT IS MORE LIKELY TO OCCUR THE LONGER THE IMPLANT IS IMPLANTED. CREATING WRINKLES OR FOLDS SHOULD BE AVOIDED DURING IMPLANTATION OR OTHER PROCEDURES AS IT CAN RESULT IN A RUPTURE IN A LATER TIME. BREAST IMPLANTS MAY ALSO SIMPLY WEAR OUT OVER TIME. MENTOR CONCLUDED THAT THE CAPSULAR CONTRACTURE IN THE PATIENT´S BREAST WAS THE RESULT OF THE BODY´S INDIVIDUAL PHYSIOLOGICAL RESPONSE TO THE IMPLANTATION OF A FOREIGN OBJECT IN SOFT TISSUE. CAPSULAR CONTRACTURE IS A KNOWN COMPLICATION ASSOCIATED WITH THESE DEVICES AND IS REFERENCED IN OUR CURRENT PRODUCT INSERT DATA SHEET. THE AMERICAN SOCIETY OF PLASTIC SURGEONS RECOMMENDS AND ENCOURAGES MEMBER SURGEONS TO ALWAYS SUBMIT BREAST IMPLANTS, CAPSULE, AND EFFUSION TO PATHOLOGY FOR EXAMINATION. GRANULOMAS ARE NONCANCEROUS NODULES THAT ARE A COMMON TISSUE REACTION TO THE PRESENCE OF A VARIETY OF FOREIGN MATERIALS, SUCH AS SILICONE. AS PART OF OUR QUALITY PROCESS, THE MANUFACTURING RECORDS OF LOT 6925217 WERE REVIEWED AND THE MANUFACTURING STANDARDS WERE MET PRIOR TO THE RELEASE OF THIS LOT. EACH DEVICE IS VISUALLY INSPECTED DURING MANUFACTURING TO ENSURE THE DEVICE MEETS THE REQUIRED SPECIFICATIONS PRIOR TO SHIPMENT. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Additional Manufacturer Narrative · 0

ON (B)(6) 2020, MENTOR BECAME AWARE THAT THE CORRECT DATE OF IMPLANT EXPLANTATION SURGERY WAS (B)(6) 2020. MENTOR ALSO BECAME AWARE THAT THE PATIENT ALSO EXPERIENCED RIGHT BREAST CAPSULAR CONTRACTURE; BAKER GRADE UNKNOWN AND RIGHT SIDE GRANULOMA. IN ADDITION, THE PATIENT ALSO REPORTED UNSPECIFIED SYSTEMIC SYMPTOMS AND WAS CONCERNED ABOUT BREAST IMPLANT ILLNESS. IMAGING REPORTS OF THE PATIENT'S BREAST FOUND THAT THE RIGHT AXILLARY LYMPH NODE DEMONSTRATES SONOGRAPHIC CHARACTERISTICS OF TYPICAL OF SILICONE UPTAKE. ULTRASOUND REPORT SUGGESTED OF SILICONE LYMPHADENOPATHY IN THE RIGHT AXILLA. ON (B)(6) 2020, THE MENTOR FAILURE ANALYSIS LAB HAS RECEIVED THE DEVICE FOR EVALUATION. THE ANALYSIS HAS BEGUN BUT IS NOT COMPLETE AT THIS TIME. WHEN THE INVESTIGATIONAL ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. REASON FOR DEVICE EXPLANT AND/OR REOPERATION: MATERIAL RUPTURE, CAPSULAR CONTRACTURE, GENERALIZED ILLNESS, GRANULOMA MENTOR IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH MENTOR HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, MENTOR, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, MENTOR, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF CERTAIN INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Additional Manufacturer Narrative · 0

ON (B)(6) 2020, MENTOR BECAME AWARE THAT THE PATIENT WAS SCHEDULED FOR IMPLANT EXPLANTATION IN (B)(6) 2020. SINCE THE DEVICE HAS NOT BEEN RETURNED FOR ANALYSIS, NO PRODUCT FAILURE ANALYSIS CAN BE CONDUCTED, AND NO DETERMINATION OF POSSIBLE CONTRIBUTING FACTORS CAN BE MADE. AS SUCH, THE INVESTIGATION WILL BE CLOSED. IF THE COMPLAINT DEVICE IS RECEIVED IN THE FUTURE, THE INVESTIGATION WILL BE REOPENED AND CONDUCTED AS APPROPRIATE. REASON FOR DEVICE EXPLANT AND/OR REOPERATION: MATERIAL RUPTURE MENTOR IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH MENTOR HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, MENTOR, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, MENTOR, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF CERTAIN INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Additional Manufacturer Narrative · 1

AT THE TIME OF THIS REPORT, MENTOR HAS RECEIVED NO INFORMATION REGARDING EXPLANTATION OR AN EXPECTED EXPLANTATION DATE. IT IS UNKNOWN AT THIS TIME IF THE DEVICE WILL BE MADE AVAILABLE FOR RETURN. AS A RESULT, NO PRODUCT FAILURE ANALYSIS CAN BE CONDUCTED, AND NO DETERMINATION OF POSSIBLE CONTRIBUTING FACTORS CAN BE MADE. AS SUCH, THE INVESTIGATION WILL BE CLOSED. IF THE COMPLAINT DEVICE IS RECEIVED IN THE FUTURE, THE INVESTIGATION WILL BE REOPENED AND CONDUCTED AS APPROPRIATE. A MANUFACTURING RECORD EVALUATION (MRE) WAS PERFORMED, AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE MRE VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. REASON FOR DEVICE EXPLANT AND/OR REOPERATION: N/A. MENTOR IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH MENTOR HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, MENTOR, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, MENTOR, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF CERTAIN INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT A (B)(6) YEAR OLD FEMALE PATIENT WHO UNDERWENT PRIMARY BREAST AUGMENTATION WITH TWO 275CC MENTOR MEMORYGEL BREAST IMPLANTS, SUFFERED RIGHT BREAST IMPLANT RUPTURE, POST-OPERATIVELY. ULTRASOUND ON (B)(6) 2020 AND MRI ON (B)(6) 2020 CONFIRMED THE RUPTURE. AT THE TIME OF THIS REPORT, MENTOR HAS RECEIVED NO INFORMATION REGARDING EXPLANTATION OR AN EXPECTED EXPLANTATION DATE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
475209 MENTOR MEMORYGEL BREAST IMPLANT PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED FTR MENTOR TEXAS 3502751BC 6925217 00081317000273

Patients

Seq Age Sex Outcome Treatment
1 44 YR Other| R