FDA Adverse Event Malfunction Summary report: N

CONTOUR TEST STRIPS (50)

MDR report key: 1001213 · Received February 15, 2008

Report

Report Number
1826988-2008-00230
Event Type
Malfunction
Date Received
February 15, 2008
Date of Event
February 4, 2008
Report Date
February 4, 2008
Manufacturer
BAYER HEALTHCARE, LLC
Product Code
NBW
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE QA LAB FOUND THE RETURNED REAGENT TO READ AN AVERAGE OF 4 MG/DL HIGH, OUT OF SPECIFICATION. PERFORMANCE WAS SATISFACTORY USING IQA RETENTION REAGENT.

Description of Event or Problem · 1

THE CUSTOMER STATED THAT HE OPENED A NEW CARTON OF TEST STRIPS AND FOUND THE CAP OPEN ON ONE OF THE BOTTLES. NO ADVERSE EVENTS WERE ALLEGED. THE TEST STRIPS WERE RETURNED FOR EVAL, AS EXPOSURE TO MOISTURE CAN CAUSE HIGH TEST RESULTS. REPLACEMENT TEST STRIPS WERE SENT TO THE CUSTOMER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CONTOUR TEST STRIPS (50) BLOOD GLUCOSE TEST STRIPS NBW BAYER HEALTHCARE, LLC 7080D 7FC3A02

Patients

Seq Age Sex Outcome Treatment
1 UNK