CONTOUR TEST STRIPS (50)
MDR report key: 1001184
·
Received February 15, 2008
Report
- Report Number
- 1826988-2008-00216
- Event Type
- Malfunction
- Date Received
- February 15, 2008
- Date of Event
- February 5, 2008
- Report Date
- February 5, 2008
- Manufacturer
- BAYER HEALTHCARE, LLC
- Product Code
- NBW
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
THE CUSTOMER STATED THAT HE TESTED HIS BLOOD GLUCOSE AND RECEIVED A READING OF 570 MG/DL USING HIS CONTOUR METER. HE RETESTED USING ANOTHER METER AND RECEIVED A READING OF 240 MG/DL. THE DIFFERENCE BETWEEN THE READINGS FALLS IN THE "C" ZONE OF THE PARKES ERROR GRID, MAKING THE DIFFERENCE CLINICALLY SIGNIFICANT. NO ADVERSE EVENTS WERE ALLEGED. THE TEST STRIPS ARE TO BE RETURNED FOR EVAL. REPLACEMENT TEST STRIPS WERE SENT TO THE CUSTOMER.