FDA Adverse Event Malfunction Summary report: N

ASCENSIA CONTOUR

MDR report key: 1001167 · Received February 15, 2008

Report

Report Number
1826988-2008-00194
Event Type
Malfunction
Date Received
February 15, 2008
Date of Event
January 30, 2008
Report Date
January 30, 2008
Manufacturer
BAYER HEALTHCARE, LLC
Product Code
NBW
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

THE CUSTOMER TESTED HER BLOOD GLUCOSE USING HER 2 CONTOUR METERS. THE NEW CONTOUR METER READ 269 MG/DL, WHILE THE OLDER METER READ 159 MG/DL. THE DIFFERENCE BETWEEN THE READINGS FALLS IN THE "C" ZONE OF THE PARKES ERROR GRID, MAKING THE DIFFERENCE CLINICALLY SIGNIFICANT. NO ADVERSE EVENTS WERE ALLEGED. THE TEST STRIPS ARE TO BE RETURNED FOR EVAL. REPLACEMENT TEST STRIPS WERE SENT TO THE CUSTOMER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ASCENSIA CONTOUR BLOOD GLUCOSE METER NBW BAYER HEALTHCARE, LLC 7151B NA

Patients

Seq Age Sex Outcome Treatment
1 UNK