CONTOUR TEST STRIPS (50)

MDR report key: 1001149 · Received February 15, 2008

Report

Report Number
1826988-2008-00181
Event Type
Malfunction
Date Received
February 15, 2008
Date of Event
January 28, 2008
Report Date
January 28, 2008
Manufacturer
BAYER HEALTHCARE, LLC
Product Code
NBW
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

THE CUSTOMER STATED THAT HIS BLOOD GLUCOSE READINGS HAD BEEN HIGHER WHEN TESTING WITH HIS CONTOUR METER. HE PERFORMED CONTROL TESTS WHILE TROUBLESHOOTING AND RECEIVED A RESULT OF 221 MG/DL. THE NORMAL CONTROL RANGE WAS 103-143 MG/DL. NO ADVERSE EVENTS WERE ALLEGED. THE TEST STRIPS WERE RETURNED FOR EVALUATION. REPLACEMENT TEST STRIPS WERE SENT TO THE CUSTOMER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CONTOUR TEST STRIPS (50) BLOOD GLUCOSE TEST STRIPS NBW BAYER HEALTHCARE, LLC NOT PROVIDED 7JC3A01