FDA Adverse Event Malfunction Summary report: N

MINOS MEDIUM BUR GUARD

MDR report key: 1001068 · Received February 14, 2008

Report

Report Number
1017294-2008-00108
Event Type
Malfunction
Date Received
February 14, 2008
Report Date
January 16, 2008
Manufacturer
CONMED LINVATEC
Product Code
HTT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION FINDINGS: CONMED LINVATEC RECEIVED THE BUR GUARD FOR EVALUATION AND CONFIRMED OVERHEATING RELATED TO BEARING ASSEMBLY FAILURE. A REVIEW OF REPAIR HISTORY RECORDS SHOW THAT THIS BUR GUARD WAS LAST REPAIRED IN 2005. THE INSTRUCTION MANUAL INFORMS THE USER TO INSPECT AND OPERATE THIS DEVICE FOR PROPER FUNCTION PRIOR TO USE. THE FOLLOWING WARNING IS PROVIDED IN THE MANUAL: DO NOT USE AIR DRILL WITHOUT THE PROPER BUR GUARD. ALSO, BE SURE THE ATTACHMENT IS PROPERLY LUBRICATED AND THE BEARING IS FUNCTIONING. FROZEN BEARINGS WILL CAUSE HEAT BUILD-UP AND POSSIBLE BURNS. THE CUSTOMER WILL BE CONTACTED WITH ADDITIONAL INFO REGARDING THIS DEVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT DURING USE OF THIS BUR GUARD IN ORAL SURGERY, IT STARTED TO SMOKE AND BECAME HOT TO THE TOUCH. THE SURGEON REPLACED THE HANDPIECE AND BUR GUARD WITH AN ALTERNATE AND COMPLETED THE PROCEDURE WITHOUT INJURY OR SURGICAL DELAY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MINOS MEDIUM BUR GUARD BUR GUARD HTT CONMED LINVATEC NA NA

Patients

Seq Age Sex Outcome Treatment
1 NK NK