FDA Adverse Event
Injury
Summary report: N
GENTLEPOWER LUX CONTRA ANGLE 25LPA
MDR report key: 1001059
·
Received February 22, 2008
Report
- Report Number
- 8010493-2008-00004
- Event Type
- Injury
- Date Received
- February 22, 2008
- Report Date
- February 5, 2008
- Manufacturer
- KALTENBACH & VOIGT GMBH
- Product Code
- EKK
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OK, US
- Reporter Occupation
- DENTIST
Narratives
Additional Manufacturer Narrative · 1
PATIENT WAS PRESCRIBED 800 MG IBUPROFEN. DUE TO IMPROPER MAINTENANCE, BEARINGS WERE WORN AND GRITTY IN DRIVE ASSEMBLY AND SHAFT. BEARINGS FALLING APART. MEDIUM DEBRIS LEVEL. PROPER MAINTENANCE TO HANDPIECE WAS DISCUSSED. REPLACEMENT WAS SENT.
Description of Event or Problem · 1
BACKCAP OF THE HANDPIECE OVERHEATED AND CAUSED A SMALL BURN TO PATIENT'S LOWER LIP.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | GENTLEPOWER LUX CONTRA ANGLE 25LPA | DENTAL HANDPIECES AND ACCESSORIES | EKK | KALTENBACH & VOIGT GMBH | 25LPA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | YR | Required Intervention |