SYRINGE 50ML LL PRECISE
Report
- Report Number
- 8041187-2020-00228
- Event Type
- Malfunction
- Date Received
- April 29, 2020
- Date of Event
- April 7, 2020
- Report Date
- June 9, 2020
- Manufacturer
- BECTON DICKINSON MEDICAL (SINGAPORE)
- Product Code
- FMF
- PMA / PMN Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SN
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THE FOLLOWING FIELDS WERE UPDATED DUE TO ADDITIONAL INFORMATION: D.10 DEVICE AVAILABLE FOR EVAL YES, RETURNED TO MANUFACTURER ON: 2020-05-27. H.6. INVESTIGATION SUMMARY ONE PHOTO AND ONE SAMPLE WERE RECEIVED BY OUR QUALITY TEAM FOR EVALUATION. UPON VISUAL INSPECTION IT WAS OBSERVED THAT THE TIP HAS A SLANT; THEREFORE, THE INCIDENT COULD BE VERIFIED. A REVIEW OF THE INTERNAL MANUFACTURING DEVICE RECORDS AND RAW MATERIAL HISTORY FILES FOR THE REPORTED LOT NUMBERS WAS PERFORMED AND NO RECORDED QUALITY PROBLEMS OR REJECTIONS TO THIS INCIDENT WERE FOUND. AN INVESTIGATION WAS INITIATED BY CONDUCTING A REVIEW OF THE BATCH RECORD FOR AFFECTED BATCH. THE REVIEW WAS FOCUSED ON THE MOLDING OPERATION AND ASSEMBLY OPERATION. THERE IS AN OVERHEAD CONVEYOR TRANSFER MOLDED BARREL FROM MOLDING OPERATION TO ASSEMBLY OPERATION. DURING THE TRANSFER, THE PRODUCT COULD HAVE GOTTEN STUCK FROM THE OVERHEAD CONVEYOR GAP RESULTING IN PRESSING AGAINST THE BARREL TIP CAUSING THE OBSERVED SLANT. THIS INCIDENT HAS BEEN ADDED TO OUR DATABASE OF REPORTED INCIDENTS. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.
IT WAS REPORTED THAT SYRINGE 50ML LL PRECISE HAD A BENT TIP. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "NOZZLE SLANTED".
DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED THAT SYRINGE 50ML LL PRECISE HAD A BENT TIP. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "NOZZLE SLANTED".
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 476194 | SYRINGE 50ML LL PRECISE | SYRINGE | FMF | BECTON DICKINSON MEDICAL (SINGAPORE) | 9353749 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |