DIMENSION CLINICAL CHEMISTRY SYSTEM
Report
- Report Number
- 1226181-2008-00015
- Event Type
- Malfunction
- Date Received
- February 19, 2008
- Date of Event
- January 21, 2008
- Report Date
- January 21, 2008
- Manufacturer
- SIEMENS HEALTHCARE DIAGNOSITICS INC.
- Product Code
- MMI
- PMA / PMN Number
- K010061
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- OTHER
Narratives
NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED. ANALYSIS OF THE INSTRUMENT AND INSTRUMENT DATA INDICATE THAT THE CAUSE FOR THE FALSELY ELEVATED TROPONIN I RESULT WAS AN OVERACTIVE R2 TRANSDUCER. A SIEMENS HEALTHCARE DIAGNOSTICS INC FIELD SERVICE REP WAS DISPATCHED TO THE ACCOUNT AND CORRECTED THE MALFUNCTION. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS.
A FALSELY ELEVATED TROPONIN I RESULT OF 1.28 NG/ML WAS OBTAINED ON A PT SAMPLE. THE RESULT WAS NOT REPORTED TO THE PHYSICIAN. THE SAMPLE WAS RETESTED AND RESULT OF 0.02 NG/ML WAS OBTAINED. PT TREATMENT WAS NOT PRESCRIBED OR ALTERED. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE FALSELY ELEVATED TROPONIN I RESULT. ANALYSIS OF THE INSTRUMENT AND INSTRUMENT DATA INDICATE THAT THE CAUSE FOR THE FALSELY ELEVATED TROPONIN RESULT WAS AN OVERACTIVE R2 TRANSDUCER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DIMENSION CLINICAL CHEMISTRY SYSTEM | CLINICAL CHEMISTRY SYSTEM | MMI | SIEMENS HEALTHCARE DIAGNOSITICS INC. | XPANDPLUSHM | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |