FDA Adverse Event
Malfunction
Summary report: N
9900
MDR report key: 1000582
·
Received February 21, 2008
Report
- Report Number
- 1720753-2008-16328
- Event Type
- Malfunction
- Date Received
- February 21, 2008
- Date of Event
- February 14, 2008
- Report Date
- February 21, 2008
- Manufacturer
- GE OEC MEDICAL SYSTEMS, INC
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NE, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SVC REP PERFORMED AN ON SITE INVESTIGATION. THE MALFUNCTION WAS VERIFIED. REPLACED THE FLUORO FUNCTION BOARD AND RELOADED THE CAL FILES. THE SYS WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PLACED BACK INTO SVC.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE IRIS CLOSED DOWN ON THE 9900 SYS. IT WAS ALSO NOTED THAT THE SYS PRESENTED A COLLIMATOR CAL NEEDED MESSAGE. THERE WAS NO REPORT OF PT INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 9900 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS, INC | 9900 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|