Manufacturer
ABBOTT DIABETES CARE LIMITED UK
Product Code
NBW
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE METER DID NOT CONFIRM FOR INCORRECT DATE AND TIME. HOWEVER, THERE IS A KNOWN MALFUNCTION WITH THE PRECISION LINK SOFTWARE THAT CAN LEAD TO INCORRECT DATE AND TIME ISSUES WHEN THE DATA IS UPLOADED TO A COMPUTER. THIS OCCURS WHEN RESULTS OBTAINED ON A METER WITH INCORRECT DATE AND TIME ARE UPLOADED TO PRECISION LINK. CUSTOMERS AND RETAILERS HAVE BEEN NOTIFIED.

Description of Event or Problem · 1

A CUSTOMER REPORTED RECEIVING AN E6 MESSAGE ON THE DISPLAY OF THEIR PRECISION XTRA BLOOD GLUCOSE METER. IT WAS THEN ADDITIONALLY IDENTIFIED BY ADC CUSTOMER SERVICE THAT THE DATE AND TIME SETTINGS IN THEIR METER WERE NOT PROPERLY SET, AND THEY REPORTED TO BE A USER OF THE PRECISION LINK DATA MANAGEMENT SYSTEM. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY OR MISTREATMENT ASSOCIATED WITH THIS EVENT.