ICL (IMPLANTABLE COLLAMER LENS)
Report
- Report Number
- 2023826-2008-00215
- Event Type
- Injury
- Date Received
- February 21, 2008
- Date of Event
- December 20, 2007
- Report Date
- January 28, 2008
- Manufacturer
- STAAR SURGICAL CO.
- Product Code
- HQL
- PMA / PMN Number
- P030016
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SP
- Reporter Occupation
- PHYSICIAN
Narratives
CONCLUSIONS - AT THE CONCLUSION OF THE ORIGINAL INVESTIGATION OPENED FOR THE ISSUE OF ICL VAULTING (BOTH INADEQUATE AND EXCESSIVE), IT WAS DETERMINED THAT THE MOST LIKELY ROOT CAUSE FOR BOTH INADEQUATE AND EXCESSIVE VAULT IS DIFFICULTY IN CLINICAL SIZING. IN NO CASE HAS A RETURNED LENS EXHIBITED A LENGTH MEASUREMENT OUTSIDE THE EXPECTED RANGE ACCORDING TO THE LABELED LENGTH. PRE-OPERATIVE MEASUREMENT TECHNOLOGY AVAILABLE TO CLINICIANS IN THE PAST DID NOT PROVIDE A DIRECT MEASUREMENT OF THE SULCUS, TO WHICH THE ICL SHOULD BE SIZED. THE CURRENT PRACTICE IN SELECTING AN APPROPRIATE ICL FOR A PATIENT IS TO MEASURE WHITE-TO-WHITE AND ANTERIOR CHAMBER DEPTH AND, THROUGH CORRELATIVE ALGORITHMS, PREDICT THE LENGTH OF THE ICL THAT WILL BRIDGE THE SULCUS. THIS METHOD OF SIZING BASED UPON WHITE-TO-WHITE AND ANTERIOR CHAMBER DEPTH MEASUREMENTS WAS USED IN THE FDA CLINICAL TRIAL AND IS RECOMMENDED IN THE CURRENT LABELING OF THE ICL. IF THE PREDICTED LENGTH IS TOO SHORT, THE RESULT MAY BE AN INADEQUATE VAULT. IF THE PREDICTED LENGTH IS TOO LONG, THE RESULT MAY BE AN EXCESSIVE VAULT. EVALUATION OF NEWLY AVAILABLE, ALTERNATE MEASURING DEVICES IS ONGOING.
IT WAS REPORTED THAT THE SURGEON INSERTED AN ICM 120V4 12.0MM IMPLANTABLE COLLAMER LENS IN 2007. DURING THE PT'S LAST VISIT, THE SURGEON DISCOVERED INADEQUATE VAULTING OF THE LENS AND THE PT WAS EXPERIENCING INTENSE PHOTOPHOBIA. THE LENS REMAINS IMPLANTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ICL (IMPLANTABLE COLLAMER LENS) | INTRAOCULAR LENS | HQL | STAAR SURGICAL CO. | ICM120V4 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 45 YR | FOAM TIP PLUNGER: MODEL, LOT NUMBER UNK| CARTRIDGE: MODEL UNK, LOT NUMBER UNK| INJECTOR: MODEL UNK, LOT NUMBER UNK |