FDA Adverse Event Malfunction Summary report: N

BLADE MERLIN, CUDA, 15 DEGREES CURVED 4.2MM

MDR report key: 1000254 · Received February 14, 2008

Report

Report Number
1017294-2008-00147
Event Type
Malfunction
Date Received
February 14, 2008
Date of Event
January 17, 2008
Report Date
January 17, 2008
Manufacturer
CONMED LINVATEC
Product Code
GFA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION FINDINGS: AN INVESTIGATION COULD NOT BE PERFORMED BECAUSE THE DEVICE WAS NOT RETURNED TO THE MFR. A REVIEW OF PRODUCT HISTORY SHOWS OTHER SIMILAR REPORTS FOR METAL SHAVINGS. THIS TYPE OF PROBLEM CAN OCCUR IF EXCESSIVE LATERAL FORCE IS APPLIED DURING USE OR IF THERE IS INSUFFICIENT LUBRICATION BETWEEN THE INNER AND OUTER BLADE. THE EXCESSIVE FORCE CAN CAUSE FRICTION AND WEAR ON THE BLADE. THE CUSTOMER WILL BE CONTACTED WITH ADDITIONAL INFO REGARDING THIS DEVICE.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING USE OF THIS SHAVER BLADE IN A KNEE ARTHROSCOPY PROCEDURE, METAL SHAVINGS WERE OBSERVED IN THE JOINT. THE SURGEON FLUSHED THE KNEE, HOWEVER, IT IS UNK IF ALL METAL SHAVINGS WERE REMOVED FROM THE JOINT. THERE WAS NO REPORT OF PT INJURY WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 BLADE MERLIN, CUDA, 15 DEGREES CURVED 4.2MM CUDA, CURVED 4.2 MM GFA CONMED LINVATEC NA BBD42622

Patients

Seq Age Sex Outcome Treatment
1 UNK