BLADE MERLIN, CUDA, 15 DEGREES CURVED 4.2MM
Report
- Report Number
- 1017294-2008-00147
- Event Type
- Malfunction
- Date Received
- February 14, 2008
- Date of Event
- January 17, 2008
- Report Date
- January 17, 2008
- Manufacturer
- CONMED LINVATEC
- Product Code
- GFA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO, US
- Reporter Occupation
- OTHER
Narratives
INVESTIGATION FINDINGS: AN INVESTIGATION COULD NOT BE PERFORMED BECAUSE THE DEVICE WAS NOT RETURNED TO THE MFR. A REVIEW OF PRODUCT HISTORY SHOWS OTHER SIMILAR REPORTS FOR METAL SHAVINGS. THIS TYPE OF PROBLEM CAN OCCUR IF EXCESSIVE LATERAL FORCE IS APPLIED DURING USE OR IF THERE IS INSUFFICIENT LUBRICATION BETWEEN THE INNER AND OUTER BLADE. THE EXCESSIVE FORCE CAN CAUSE FRICTION AND WEAR ON THE BLADE. THE CUSTOMER WILL BE CONTACTED WITH ADDITIONAL INFO REGARDING THIS DEVICE.
IT WAS REPORTED THAT DURING USE OF THIS SHAVER BLADE IN A KNEE ARTHROSCOPY PROCEDURE, METAL SHAVINGS WERE OBSERVED IN THE JOINT. THE SURGEON FLUSHED THE KNEE, HOWEVER, IT IS UNK IF ALL METAL SHAVINGS WERE REMOVED FROM THE JOINT. THERE WAS NO REPORT OF PT INJURY WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | BLADE MERLIN, CUDA, 15 DEGREES CURVED 4.2MM | CUDA, CURVED 4.2 MM | GFA | CONMED LINVATEC | NA | BBD42622 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |