ANIMAS INSULIN INFUSION PUMP
Report
- Report Number
- 2531779-2008-00115
- Event Type
- Injury
- Date Received
- February 19, 2008
- Date of Event
- January 27, 2008
- Report Date
- January 27, 2008
- Manufacturer
- ANIMAS CORP.
- Product Code
- LZG
- PMA / PMN Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- UNKNOWN
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. THE PT STATED HE MAY HAVE INADVERTENTLY ADMINISTERED EXCESS INSULIN BY PRIMING THE PUMP WHILE ATTACHED TO THE INFUSION SET. THE PUMP USER GUIDE INSTRUCTS USERS TO DISCONNECT FROM THE INFUSION SET PRIOR TO PRIMING IT. ADDITIONALLY, THE PUMP NOTIFIES THE USER TO DISCONNECT FROM THE INFUSION SET THREE TIMES DURING THE REWIND/PRIME SEQUENCE. THE PT HAS CONTINUED TO USE THE PUMP WITH NO REPORTED SUBSEQUENT EVENTS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
THE PT RECEIVED EMS TREATMENT FOR HYPOGLYCEMIA AND INEBRIATION. THE PT STATED THAT HE MAY HAVE ADMINISTERED EXCESS INSULIN BY PRIMING THE PUMP WHILE ATTACHED TO THE INFUSION SET.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANIMAS INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | ANIMAS CORP. | ANIMAS 2020 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | YR | Required Intervention |