FDA Adverse Event
Malfunction
Summary report: N
YELLOFINS STIRRUP
MDR report key: 10000438
·
Received April 27, 2020
Report
- Report Number
- 10000438
- Event Type
- Malfunction
- Date Received
- April 27, 2020
- Date of Event
- November 22, 2019
- Report Date
- April 21, 2020
- Manufacturer
- ALLEN MEDICAL SYSTEMS, INC.
- Product Code
- CCX
- Product Problem
- Yes
- Report Source
- User Facility report
- Reporter Location
- FL
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Description of Event or Problem · 1
RIGHT LEG STIRRUP CAME OUT OF BRACKET WITH PATIENT'S LEG IN THE STIRRUP. THE STIRRUP CAME OUT OF THE STIRRUP HOLDER DURING SURGERY. SURGEON WAS STANDING NEXT TO LEG AND WAS ABLE TO HOLD LEG AND IT WAS AT THE END OF THE CASE. NO HARM TO THE PATIENT. BIOMED TECH TESTED YELLOFINS STIRRUPS. CONNECTED BOTH BRACKETS TO SURGERY BED. BRACKETS TIGHT AND STIRRUP WAS MOVED TO DIFFERENT DIRECTIONS. UNIT WAS STEADY WITH NO ISSUES. STIRRUPS RETURNED TO SERVICE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 464230 | YELLOFINS STIRRUP | SUPPORT, PATIENT POSITION | CCX | ALLEN MEDICAL SYSTEMS, INC. | O-YFASI |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 16060 DA |