FDA Adverse Event Malfunction Summary report: N

YELLOFINS STIRRUP

MDR report key: 10000438 · Received April 27, 2020

Report

Report Number
10000438
Event Type
Malfunction
Date Received
April 27, 2020
Date of Event
November 22, 2019
Report Date
April 21, 2020
Manufacturer
ALLEN MEDICAL SYSTEMS, INC.
Product Code
CCX
Product Problem
Yes
Report Source
User Facility report
Reporter Location
FL
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Description of Event or Problem · 1

RIGHT LEG STIRRUP CAME OUT OF BRACKET WITH PATIENT'S LEG IN THE STIRRUP. THE STIRRUP CAME OUT OF THE STIRRUP HOLDER DURING SURGERY. SURGEON WAS STANDING NEXT TO LEG AND WAS ABLE TO HOLD LEG AND IT WAS AT THE END OF THE CASE. NO HARM TO THE PATIENT. BIOMED TECH TESTED YELLOFINS STIRRUPS. CONNECTED BOTH BRACKETS TO SURGERY BED. BRACKETS TIGHT AND STIRRUP WAS MOVED TO DIFFERENT DIRECTIONS. UNIT WAS STEADY WITH NO ISSUES. STIRRUPS RETURNED TO SERVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
464230 YELLOFINS STIRRUP SUPPORT, PATIENT POSITION CCX ALLEN MEDICAL SYSTEMS, INC. O-YFASI

Patients

Seq Age Sex Outcome Treatment
1 16060 DA