Date
Recalling firm
Status
Reason
2026-05-04
Open, Classified
During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
2026-04-27
Open, Classified
Kits contain recalled lot of Ferndale Laboratories MASTISOL Liquid Adhesive. Butyrate tubing may crack during actuation and render product unusable for application.
2026-03-19
Open, Classified
During an internal review, Medline identified higher-than-expected endotoxin levels on a representative Medline neuro sponge product, indicating that in-market product may be out-of-specification for endotoxin.
2026-02-25
Open, Classified
The 510(k) regulatory clearances for certain Control Syringes, Guidewires, and/or High-Pressure Tubing have been rescinded.
2026-01-07
Open, Classified
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
2026-01-07
Open, Classified
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
2025-12-16
Open, Classified
Medline kits containing Stryker's Strykeflow 2 Suction Irrigators due to a change in design in 2024, irrigation solution is able to travel to the handpiece and battery pack during use of the Strykeflow 2 Suction Irrigator. This design change has caused some devices to leak and/or emit vaporized saline (which can resemble smoke) during use.
2025-09-30
Open, Classified
Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing SafeAir Smoke Evacuation Pencils. Stryker Instruments has expanded the scope of their recall of SafeAir Smoke Evacuation Pencils previously initiated in October 2024, adding an additional 47 lots to the scope of the recall. The initial recall was due to the potential for product to activate without manual input when plugged into a power source or remain active after buttons are released.
2025-06-11
Open, Classified
CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration.
2025-06-11
Open, Classified
CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration.
2025-06-11
Open, Classified
CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration.
2025-06-11
Open, Classified
CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration.
2025-06-11
Open, Classified
CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration.
2025-03-03
Open, Classified
There is a potential for open/weak seals on purchased breather pouches used for packaging of select sterile convenience kits.
2025-02-11
Open, Classified
Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/other quality issues have been identified that may pose a risk to patient health.
2024-11-21
Open, Classified
Custom convenience kits are being recall because the contain non-sterile Codman Surgical Patties & Surgical Strips, which have the potential for endotoxin contamination.
2024-10-09
Open, Classified
Medline is recalling medical convenience manufactured using certain lots of SafeAir Smoke Evacuation Pencil which were subsequently recalled by Stryker.
2024-09-20
Open, Classified
Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or possible foreign body reactions
2024-09-20
Open, Classified
Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or possible foreign body reactions
2024-09-20
Open, Classified
Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or possible foreign body reactions
2024-09-20
Open, Classified
Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or possible foreign body reactions
2024-09-20
Open, Classified
Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or possible foreign body reactions
2024-09-20
Open, Classified
Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or possible foreign body reactions
2024-09-20
Open, Classified
Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or possible foreign body reactions
2024-09-20
Open, Classified
Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or possible foreign body reactions
2024-09-20
Open, Classified
Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or possible foreign body reactions
2024-09-04
Open, Classified
It has been determined that certain Convenience kits that were packaged in breather pouch lot 323080002 may result in the potential of using non-sterile product if the vendor seal of the pouch is opened/unsealed.
2024-09-04
Open, Classified
It has been determined that certain Convenience kits that were packaged in breather pouch lot 323080002 may result in the potential of using non-sterile product if the vendor seal of the pouch is opened/unsealed.