Date
Recalling firm
Status
Reason
2026-05-04
Open, Classified
During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
2026-04-27
Open, Classified
Kits contain recalled lot of Ferndale Laboratories MASTISOL Liquid Adhesive. Butyrate tubing may crack during actuation and render product unusable for application.
2026-04-10
Open, Classified
Specific Medline Kits contain Cardinal Health Monoject 1mL Luer Lock Syringes, Tuberculin Print, Sterile. Cardinal Health has recalled these syringes because the outer carton and blister pack are labeled as a 1 mL Luer Lock Tuberculin Syringe (Product Code 1180100777), but the syringes inside are U-100 insulin syringes.
2026-03-19
Open, Classified
During an internal review, Medline identified higher-than-expected endotoxin levels on a representative Medline neuro sponge product, indicating that in-market product may be out-of-specification for endotoxin.
2026-02-27
Open, Classified
Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.
2026-02-27
Open, Classified
Unapproved design changes to the products outside of the 510(k) clearance.
2026-01-07
Open, Classified
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
2026-01-07
Open, Classified
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
2025-10-03
Open, Classified
Medline has identified specific sterile eye procedure kits with pouch packaging that may exhibit open seals.
2025-09-30
Open, Classified
Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing SafeAir Smoke Evacuation Pencils. Stryker Instruments has expanded the scope of their recall of SafeAir Smoke Evacuation Pencils previously initiated in October 2024, adding an additional 47 lots to the scope of the recall. The initial recall was due to the potential for product to activate without manual input when plugged into a power source or remain active after buttons are released.
2025-09-04
Open, Classified
Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.
2025-05-23
Open, Classified
Medline Industries, LP. is issuing a recall for specific Medline kits containing BD Insyte Autoguard Shielded IV Catheters 22 GA x 1.00 IN which was identified that during insertion of the Catheter, the needle is slow to retract or fails to retract.
2025-03-19
Open, Classified
It was identified that the Medline Sterile Water and Medline Sterile 0.9% Normal Saline bottles were leaking at the interface of the peel-foil and bottle opening.
2025-03-03
Open, Classified
There is a potential for open/weak seals on purchased breather pouches used for packaging of select sterile convenience kits.
2025-02-11
Open, Classified
Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/other quality issues have been identified that may pose a risk to patient health.
2025-02-11
Open, Classified
Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/other quality issues have been identified that may pose a risk to patient health.
2024-11-08
Open, Classified
Potential for endotoxin contamination in non-sterile surgical strips and patties in Presource Kits.
2024-11-08
Open, Classified
Potential endotoxin contamination in non-sterile surgical strips and patties in Presource kits.
2024-10-15
Open, Classified
Reported breakage in the tubing Anti-Reflux Valve (ARV) due to improper use. As a result, changes to labeling were made.
2024-10-09
Open, Classified
Medline is recalling medical convenience manufactured using certain lots of SafeAir Smoke Evacuation Pencil which were subsequently recalled by Stryker.
2024-09-04
Open, Classified
It has been determined that certain Convenience kits that were packaged in breather pouch lot 323080002 may result in the potential of using non-sterile product if the vendor seal of the pouch is opened/unsealed.
2024-09-04
Open, Classified
It has been determined that certain Convenience kits that were packaged in breather pouch lot 323080002 may result in the potential of using non-sterile product if the vendor seal of the pouch is opened/unsealed.