Tool · Post-Market Surveillance

PSUR Report

The vigilance database section of your PSUR, generated in seconds: FDA MAUDE adverse event trends, recalls, and enforcement classifications for your device and similar devices — with the documented, reproducible methodology auditors ask for.

Period
from
to
Leave empty for the last 24 months

Reporting period 2024-072026-06 · FDA data refreshed ~monthly

Product Code: NSA FDA class 3

Over-The-Counter Automated External Defibrillator

View full classification →
Adverse events in period
900
+603% vs. prior period (128)
Deaths reported
17
Recalls in period
0
Class I enforcement
0

Adverse events per month

2024-07 – 2026-06
Loading...

Events by type

Period vs. prior period
Loading...
Event type
Period
Prior
Death
17
4
Injury
13
0
Malfunction
870
124

Most reported coded problems

Top 15
Product problems
Count
Failure of Device to Self-Test
466
Failure to Power Up
120
Inaudible or Unclear Audible Prompt/Feedback
44
Failure to Run on Battery
40
No Audible Prompt/Feedback
39
Material Split, Cut or Torn
32
Electrical /Electronic Property Problem
22
Inappropriate Audible Prompt/Feedback
20
Key or Button Unresponsive/not Working
19
Battery Problem
16
Premature Discharge of Battery
15
Device Alarm System
15
Failure to Deliver Shock/Stimulation
14
Material Integrity Problem
8
Audible Prompt/Feedback Problem
8
Patient problems
Count
No Clinical Signs, Symptoms or Conditions
875
Cardiac Arrest
24
Insufficient Information
2
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
1

Recalls in period

0 total

No recalls recorded for these product codes in the reporting period.

Adverse events by year

Loading...

Methodology & data provenance

For your PSUR appendix

Search protocol: FDA MAUDE adverse event reports, FDA device recalls, and FDA enforcement reports (openFDA datasets) were queried for product code NSA, reporting period 2024-07 to 2026-06 (prior comparison period 2022-07 to 2024-06).

Data coverage: adverse event data for these product codes extends through 2026-05. Source openFDA export dated 2026-06-30. Recall records with missing or implausible initiation dates (0 for these codes) are excluded from period counts.

Generated: 2026-07-06 18:42 UTC by BEUDAMED (beudamed.com/fda/psur-report). Dataset sizes: 25,039,198 adverse events, 58,607 recalls, 39,365 enforcement reports.

PSUR database searches, explained

Under the EU MDR, manufacturers must produce a Periodic Safety Update Report (PSUR) (Class IIa and above; a PMS report for Class I) on a defined schedule. A required part of that report is a survey of publicly available vigilance data — most importantly the FDA's MAUDE adverse event database and FDA recall records — covering both your own device and similar devices on the market (the "state of the art").

Doing this by hand means repeated MAUDE queries, manual export, deduplication, and trending — hours of work per report, repeated every reporting period. This tool runs the same search protocol in seconds and documents it: which datasets were queried, with which filters, when, and with what coverage — so the search is reproducible, which is exactly what a Notified Body auditor wants to see.

Event counts are grouped by the FDA's reported event type (death, injury, malfunction). Name-based device matching is clearly labeled as indicative: only you can confirm whether a specific MAUDE report concerns your device. Recall records with implausible dates are excluded and the exclusion is disclosed. All data is from the official openFDA datasets and refreshed approximately monthly.